After preparing the buffer solution, it should be placed in a glass bottle or a polyethylene bottle (alkaline pH buffer solutions such as pH 9.18, pH 10.01, pH 12.46, etc. should be placed in polyethylene bottles). The bottle cap should be tightly closed and stored at low temperature (5-10 ℃) in the refrigerator. It can generally be used for about six months. If turbidity, mold or precipitation are found, it cannot be continued to be used. When using, prepare several 50mL polyethylene small bottles, pour the buffer solution from the large bottles into the small bottles, and let them stand at ambient temperature for 1-2 hours until the temperature is balanced before use. After use, it should not be poured into the large bottle to avoid contamination. The buffer solution in the bottle can be used for 2-3 days under environmental conditions of>10 stupňov. Vo všeobecnosti sa roztoky s pH 7,00, pH 6,86 a pH 14,00 môžu používať dlhší čas. Roztoky s pH 9,18 a pH 10,01 sú náchylné na zmeny pH v dôsledku absorpcie oxidu uhličitého zo vzduchu.
Aký je účel pH pufrovacieho roztoku?
Pred meraním pH kalibrujte pH meter.
Na overenie presnosti pH metra, napríklad po kalibrácii pH metra s pH 6,86 a pH 14,00, vložte pH elektródu do roztoku pH 9,18 a skontrolujte, či je zobrazená hodnota prístroja v súlade s hodnotou pH štandardného roztoku.
Je potrebné prekalibrovať pH meter počas všeobecne presného merania? Po kalibrácii a použití pH metra môže dôjsť k posunu alebo zmenám. Preto pred testovaním vložte elektródu do štandardného tlmivého roztoku, ktorý je blízko meranému roztoku, a na základe veľkosti chyby zistite, či je potrebná rekalibrácia.